Investigation 2: Spectrophotometry

Spectrophotometry finds the concentration of a colored solution by measuring how much light it absorbs, using a calibration line built from known standards to read an unknown's concentration off its absorbance.

A spectrophotometer shines light through a solution and measures how much light makes it to a detector on the other side. You set it to the wavelength the solution absorbs most strongly because that's where small changes in concentration produce the biggest, most measurable changes in absorbance. At a wavelength the solution barely absorbs, the signal barely moves even as concentration changes a lot, so your readings are noisy and imprecise.

Before measuring anything, you blank the instrument with pure solvent, the same solvent your samples sit in. The blank absorbs some light on its own, from the cuvette walls and the solvent itself. Zeroing out that background means every later reading reflects only the solute you actually care about. Then you measure a series of standards of known concentration, plot absorbance (y-axis) against concentration (x-axis), and fit a line through the points. Because Beer's Law says absorbance is directly proportional to concentration, this line is straight and passes through the origin. Its slope becomes your conversion factor: divide an unknown's absorbance by the slope and you get its concentration.

This investigation is where graphing and error analysis meet experimental design. A good calibration curve needs standards spread across a sensible range, replicate trials, and attention to anything that scatters light instead of truly absorbing it. It sits alongside the other quantitative labs in the unit as a model for turning an instrument reading into a real chemical quantity, with the line itself serving as the tool that connects data to an answer.

Key terms

Absorbance
A unitless measure of how much light a solution absorbs at a given wavelength.
Beer's Law
The relationship stating absorbance is directly proportional to the concentration of the absorbing species.
Blanking
Zeroing the spectrophotometer using pure solvent so background absorbance is not counted in later readings.
Calibration curve
A graph of absorbance versus known concentrations, used to determine an unknown concentration from its absorbance.
Cuvette
The small, clear container that holds a solution in the light path of a spectrophotometer.

Watch it explained

  1. How it worksWalk me through blanking a spectrophotometer with pure solvent before taking readings.
  2. Worked exampleGraph absorbance against concentration for five standards and use the line to find an unknown.
  3. Worked exampleFind the mass percent of copper in brass from an absorbance reading and a calibration line.
  4. Common mistakeWhy isn't a cloudy or scratched cuvette giving a valid absorbance reading?

Check yourself

Your turn1 of 2
A calibration line for a spectrophotometer has a slope of 2.24 per M (through the origin). A 1.75 g brass sample is fully dissolved into 236 mL of solution and measured to have an absorbance of 0.619. Find the mass percent of copper (63.55 g/mol) in the brass sample.